CREATING A LINEAR REGRESSION CURVE FOR LOI (LOSS OF IGNITION) TESTING PT
SEMEN TONASA HOTMEAL USING X-RAY
Grenatha Lengke1 , Muhammad Apris Cam2 ,
Rosalia Sira SarungAllo3
Indonesian Christian
University
Email: [email protected]1,
[email protected]2, [email protected]3
|
Abstract |
|
As development progresses in Indonesia, the role of
the industrial world is also experiencing rapid development. The company's
ability to produce high quality products is one of the things that can
support the company's long-term success. As the era progresses, the
application of conventional technology in the industrial world begins to
shift to an automatic model using computer devices. PT Semen Tonasa, especially in the Quality Control unit, there are
still some sample tests that are carried out manually, so it takes a long
time to carry out sample preparation to test the sample. If you calculate the
average time needed to carry out the test to test the sample, it takes 60
minutes, which is wrong. one test is the LOI (Loss Of
Ingnation) test. The aim of the research is to
create a formula that is suitable for calculating LOI (Loss of Remembrance)
specifically for hotmeal 05 and 10. The research
was carried out using direct experimental methods and data processing using
Microsoft Excel 2013 software. The research was carried out by directly
taking raw materials in the form of clinker chunks and Kiln Feed in the
factory area, then preparation and data processing were carried out. This
research was conducted at PT Semen Tonasa, Quality
Control Unit IV from April 1 2023 to June 30 2023. The results of this
research are that a linear regression equation is obtained, namely y=
1.1824x-6.9094, then this equation is input into a computer device with the
method name, namely Hotmeal_2023, to be used as a reference in working on Hotmeal samples more quickly. It is necessary to carry
out research with more samples. Again, so that the value obtained on the
computer is the same as the manual LOI calculation, and the cleanliness of
the equipment in carrying out the analysis needs to be improved so that the
sample is not contaminated by other dirt/dust. Keywords: Semen
Tonasa, Loss Of Ignition, Hotmeal. |
As development progresses in
Indonesia, the role of the industrial world is also experiencing rapid
development (Purba
et al., 2021). The company's ability to produce
high-quality products is one of the things that can support the company's
long-term success (Lasdi,
2017; Brata
& Soediantono, 2022). PT Semen Tonasa is a cement
industry that has the largest installed capacity in eastern Indonesia with an
initial design capacity of 5,980,00 tons of cement per year, but the company
has succeeded in increasing its production capabilities so that currently the
capacity has reached 6,700,000 tons of cement per year (Soamole,
2022). The manufacture of cement at PT
Semen Tonasa uses raw materials, namely clinker which is made from a mixture of
clay and limestone, limestone and other additional materials (Effendi,
2018) (Zahidin
& Rubianto, 2020). �The cement manufacturing process involves
several industrial tools, including; crusher, warehouse, belt
conveyor, silo landing, kiln, cooler, mill, etc. The economy
is currently developing very rapidly, along with the development of
increasingly sophisticated science and technology (IPTEK) (Radiansyah,
2018; Fitri
Mulyani, 2021). Thus, increasingly fierce
competition between companies encourages each company to establish control over
the quality of refrigerated production (Rasul,
2017). �As the era progresses, the application of
conventional technology in the industrial world begins to shift to an automatic
model using computer devices (Fauzi
et al., 2023). PT Semen
Tonasa, especially in the Quality Control unit, still has some sample
tests carried out manually, including the Blaine test, kiln
feed and cement residue tests, moisture, and LOI (Loss of
Ignition) so it takes a long time to carry out sample preparation to
sample testing. The average time needed to carry out the test is calculated at
60 minutes (Fajri,
2021). One type of sample that is tested
each shift is hotmeal which is the result of initial combustion at a
temperature of 530-700 o C which occurs in cylones 5 and 10. This
research aims to create a formula that is suitable for calculating LOI ( Loss
Of Ignition) specifically on hotmeal 05 and 10 (Darmalaksana,
2020). The expected benefit of the
research is for the factory where it is hoped that it will be able to provide
an overview of the formula/formula in determining LOI (Loss of Ignition)
so that it can save time and energy in carrying out sample analysis every day
at PT Semen Tonasa Quality Control, the second expected benefit is for
the author It is hoped that it will be able to increase the author's knowledge
and experience in using software in everyday life, especially in the
world of work, as well as increase the author's thinking power in finding
solutions to problems that occur.
Internship II will be held from
April 1 2023 to June 30 2023 at the Tonasa factory, Pangkep Regency, South
Sulawesi, Tonasa IV Quality Control Unit.
1.
Data collection technique
The method for carrying out data collection for this special task is,
among other things, through the interview method, namely a method carried out
by asking the resource person (operator) directly to obtain data. Literature
Study Method (literature study) is a method carried out by looking for various
references from books, journals, etc. as a basis for analysis and making
reports. The next method, namely the observation method, is a method carried
out by making direct observations and measurements in the field. The final
method, namely the discussion method, is a method that is carried out by
discussing with many people in the field the possible main causal factors in
the problem that occurs and the best way to solve the problem that can be
applied later.
2.
Working Steps
a.
Creation of standard sample series
The standard sample series was made by varying the kiln feed and
clinker to obtain an LOI range of 0.5-12%.
b.
Homogenization
The first step is to weigh 5 kg of clinker and 5 kg of kiln feed using
a container and scales, then grind the clinker to get a smaller size using a
crusher, put the sample into the vessel to get a sample in very fine
condition using a windmill. The third step, namely Put each
sample into the mixing to get a more homogeneous sample for 2 hours.
Then weigh the clinker and kiln feed samples according to the specified
proportions then combine them in one container. After that, homogenize the
mixture using a homogenizer for 30 minutes. Then put each sample into a plastic
zipper that has been named after the sample in sequence.
c.
Manual LOI Analysis
The first step is to weigh the empty Simplo and Duplo containers. Then
weigh 1 gram of each sample simply and in duplicate. Next, put the sample in the
furnace for 15 minutes, then put it in the desiccator for 10
minutes. Weigh the container + sample again using a scale. The final step is to
calculate the LOI using the formula and record the results:
d.
CaCO3 XRD analysis
The first step is to weigh 15
grams of each sample with 3 pills. Second, set the program on the HP-MP, namely
program 7 for Hotmeal. Make sure the ring is on the conveyor belt and press the
star mill and press. Then insert the sample into the HP-MP, wait until the
sample comes out of the conveyor belt and spray with air to remove dirt on the
sample that has been ground and pressed. Insert the sample container
into the XRD tool, set the program on the screen, click start on the screen and
wait until the sample has finished analyzing, then record the results.
e.
Xray Intensity Analysis
Weigh 15 grams of each sample with 3 pills, and set the program on the
HP-MP, namely program 7 for Hotmeal. Make sure the ring is on the conveyor belt
and press the star mill and press. Insert the sample into the HP-MP, wait until
the sample comes out of the conveyor belt then spray with air to remove dirt on
the sample that has been ground and pressed. Enter the sample in position
number 22 in the X-ray tool. The next step is to set it on the computer
(operational mode, click sample number 13, make sure the method "CaCO3 hot
meal" make sure the sample number is correct, change the name according to
the sample name, click start at the top), wait until the analysis results come
out and record the results of the analysis in the form kbps.
f.
Data Processing
Data processing for this special task uses Microsoft Excel, namely by
collecting data from each sample analysis in a table. Then the data is created
in graphical form so that it is easy to read and draw conclusions from each
work step carried out. The results of data processing will be further validated
through experiments.
g.
Initial Validation
The initial validation stage was carried out using the LOI equation
formula that had been obtained at the data processing stage, then testing was
carried out to ensure the accuracy of the formula in determining LOI.
h.
Evaluation result
The evaluation results were carried out when the equation was obtained,
namely y= 1.1824x-6.9094, this evaluation was carried out for 1 week (5 working
days) and the results were close to the LOI calculation value manually (gravi).
A.
Creation of Standard
Sample Series
The composition comparison between clinker and Kiln Feed
is 80%: 20%, this comparison aims to obtain an LOI range of 0.5-12%. The
creation of a standard sample series can be seen in the table below.
Table 1
Results of Making a
Standard Sample Series
|
Proportion |
LOI
HM |
|||||
|
Clinker (%) |
Sample
Weight (grams) |
Kiln
Feed (%) |
Sample
Weight (grams) |
Total
Sample Weight (grams) |
Target |
|
|
1 |
80.00 |
200.00 |
20.00 |
�������� 50.00 |
250 |
7.20 |
|
2 |
67.01 |
������ 167.52 |
32.99 |
�������� 82.48 |
12.00 |
|
|
3 |
69.60 |
������ 173.99 |
30.40 |
�������� 76.01 |
11.00 |
|
|
4 |
72.36 |
������ 180.9 |
27.64 |
�������� 69.1 |
10.00 |
|
|
5 |
75.12 |
������ 187.81 |
24.88 |
�������� 62.19 |
9.00 |
|
|
6 |
77.89 |
������ 194.72 |
22.11 |
�������� 55.28 |
8.00 |
|
|
7 |
80.65 |
������ 201.63 |
19.35 |
�������� 48.37 |
7.00 |
|
|
8 |
83.42 |
������ 208.54 |
16.58 |
������� 41.46 |
6.00 |
|
|
9 |
86.18 |
������ 215.45 |
13.82 |
�������� 34.55 |
5.00 |
|
|
10 |
88.94 |
������ 222.36 |
11.06 |
�������� 27.64 |
4.00 |
|
|
11 |
91.71 |
������ 229.27 |
8.29 |
20.73 |
3.00 |
|
|
12 |
94.47 |
236.18 |
5.53 |
13.82 |
2.00 |
|
|
13 |
97.24 |
243.09 |
2.76 |
6.91 |
1.00 |
|
|
14 |
98.62 |
246.55 |
1.38 |
3.45 |
0.50 |
|
|
Total Sample Weight |
2908.01 |
|
591.98 |
|
|
|
It can be seen that in determining the ideal sample
weight, a proportion must be used that must match the clinker content to the
Kiln Feed content. At this stage, the standard series was created 13 times to
obtain a target LOI value of 0.5-12% with the sample weight remaining the same,
namely 250 grams.
B.
Manual LOI Analysis
Manual LOI testing is carried out as a basis for
comparison between manual LOI values and LOI values using X-ray (Sentosa et al., 2018). The results of the manual LOI analysis are in the table
below
Table 2
results of manual LOI
calculations
|
No |
Proportion |
LOI HM |
|||||
|
Clinker (%) |
Sample Weight (grams) |
Kiln Feed (%) |
Sample Weight (grams) |
Total sample weight (grams) |
Target |
Actual |
|
|
1 |
80.00 |
200.00 |
20.00 |
50.00 |
250 |
7.20 |
|
|
2 |
67.01 |
167.52 |
32.99 |
82.48 |
12.00 |
12.02 |
|
|
3 |
69.60 |
173.99 |
30.40 |
76.01 |
11.00 |
11.03 |
|
|
4 |
72.36 |
180.90 |
27.64 |
69.10 |
10.00 |
10.14 |
|
|
5 |
75.12 |
187.81 |
24.88 |
62.19 |
9.00 |
9.17 |
|
|
6 |
77.89 |
194.72 |
22.11 |
55.28 |
8.00 |
8.11 |
|
|
7 |
80.65 |
201.63 |
19.35 |
48.37 |
7.00 |
7.00 |
|
|
8 |
83.42 |
208.54 |
16.58 |
41.46 |
6.00 |
5.97 |
|
|
9 |
86.18 |
215.45 |
13.82 |
34.55 |
5.00 |
5.17 |
|
|
10 |
88.94 |
222.36 |
11.06 |
27.64 |
4.00 |
4.33 |
|
|
11 |
91.71 |
229.27 |
8.29 |
20.73 |
3.00 |
3.19 |
|
|
12 |
94.47 |
236.18 |
5.53 |
13.82 |
2.00 |
2.49 |
|
|
13 |
97.24 |
243.09 |
2.76 |
6.91 |
1.00 |
1.32 |
|
|
14 |
98.62 |
246.55 |
1.38 |
3.45 |
0.50 |
0.88 |
|
�����������
From
Table 2 it can be seen that the results of manual LOI calculation have
difference slight difference between the target value with mark manual LOI
calculation. From 12 samples 2 samples have a difference of 2-3 points from the
target ie samples no. 2 and 3, meanwhile, sample number 7 results from manual
LOI calculation is the same with a target of 7, for
sample other has quite a difference Far off target, however, value still made
reference. The manual LOI calculation value
varies from the target because it is caused by different sample treatments.
C.
CaCO3 XRD analysis
The results of CaCO3 analysis using XRD spectrophotometry
can be seen in the table below.
Table 3
Results of CaCO3
analysis using an XRD tool
|
Sample |
CaCO3 value |
|
|
1 |
LOI 0.5 |
2.19 |
|
2 |
LOI 1 |
3.43 |
|
3 |
LOI 2 |
5.34 |
|
4 |
LOI 3 |
7.05 |
|
5 |
LOI 4 |
8.93 |
|
6 |
LOI 5 |
10.91 |
|
7 |
LOI 6 |
13.01 |
|
8 |
LAW 7 |
14.97 |
|
9 |
LAW 8 |
16.80 |
|
10 |
LAW 9 |
19.40 |
|
11 |
LAW 10 |
21.74 |
|
12 |
LAW 11 |
23.36 |
|
13 |
LAW 12 |
25.44 |
The results of the CaCO3 analysis used in each sample
were analyzed (SETIOSOSARI, 2018).
Figure 1
LOI graph with CaCO
content
D.
Xray Intensity
Analysis
X-ray analysis was carried out to obtain the composition
of the organic material content in the samples being analyzed (Sahdiah & Kurniawan,
2023). There were 12 samples used in this experiment and the
results are in the table below.
Table 4
�XRay intensity results
|
No |
Sample |
Intensity Xray |
|
1 |
LAW 0.5 |
0.3934 |
|
2 |
LAW 1 |
0.4058 |
|
3 |
LAW 2 |
0.4216 |
|
4 |
LAW 3 |
0.4337 |
|
5 |
LAW 4 |
0.4418 |
|
6 |
LAW 5 |
0.4750 |
|
7 |
LAW 6 |
0.4939 |
|
8 |
LAW 7 |
0.5011 |
|
9 |
LAW 8 |
0.5299 |
|
10 |
LAW 9 |
0.5377 |
|
11 |
LOI 10 |
0.5762 |
|
12 |
LOI 11 |
0.5940 |
|
13 |
LOI 12 |
0.6166 |
The calculation results of the Xray intensity analysis
are graphed with the results of the CaCO3 XRD analysis in Table 4 above, so the
results are as follows:
Figure 2
Linear graph between
intensity vs. CaCO3 XRD
The linear regressive equation in the graph above, namely
y= 104.89x-38.529, is used to calculate the estimated CaCO3 Xray value using
the resulting value of the CaCO3 Xray intensity as the X variable, the results
are in the table below.
Table 5
results in CaCO3 Xray estimation
calculation
|
CaCO3 Xray INTENSITY |
CaCO3 Xray Estimation |
|
|
LOI 0.5 |
0.6166 |
2.73 |
|
LOI 1 |
0.594 |
4.04 |
|
LOI 2 |
0.5762 |
5.69 |
|
LOI 3 |
0.5377 |
6.96 |
|
LOI 4 |
0.5299 |
7.81 |
|
LOI 5 |
0.5011 |
11.29 |
|
LOI 6 |
0.4939 |
13.28 |
|
LOI 7 |
0.475 |
14.03 |
|
LOI 8 |
0.4418 |
17.05 |
|
LOI 9 |
0.4337 |
17.87 |
|
LOI 10 |
0.4216 |
21.91 |
|
LOI 11 |
0.4058 |
23.78 |
|
LOI 12 |
0.3934 |
26.15 |
The results in the table above were then made into a
graph with the LOI gravity values, the results are as follows:
Figure 3
Linearity graph
between CaCO3 Xray vs LOI Gravi
�����������
The linear regression equation in the picture above,
namely y= 0.4785x-0.1399, is used to calculate the LOI X-Ray value with the x
value, namely CaCO3 X-Ray estimation so that the results are:
Table 6
Results of
calculating LOI XRay values
|
SAMPLE |
LOI
X-ray |
|
LOI 0.5 |
1.17 |
|
LOI 1 |
1.80 |
|
LOI 2 |
2.59 |
|
LOI 3 |
3.20 |
|
LOI 4 |
3.60 |
|
LOI 5 |
5.27 |
|
LOI 6 |
6.22 |
|
LOI 7 |
6.58 |
|
LOI 8 |
8.02 |
|
LOI 9 |
8.42 |
|
LOI 10 |
10.35 |
|
LOI 11 |
11.24 |
|
LOI 12 |
12.38 |
�����������
Next
step is to make a chart connection linearity between CaCO3 Xray intensity of
LOI Gravi so that the result that is.
Figure 4
Linear graph between
CaCO3 Xray intensity vs. LOI Gravi
�����������
After knowing the linear regression equation, the
equation that will be tested on the computer is y = 50.191x-18.572. After the
equation has been entered into the computer, the next step is to prepare the
sample for testing manually and automatically using a computer. The results are
in Table 7 below (Ariadi & Fithriasari,
2016).
Table 7
Results of LOI
analysis using a computer program and manual LOI
|
No |
CaCO3 |
LOI
man |
|
1 |
6.47 |
0.57 |
|
2 |
6.51 |
0.91 |
|
3 |
6.67 |
1.14 |
|
4 |
6.81 |
1.2 |
|
5 |
6.56 |
1.14 |
|
6 |
7.18 |
1.59 |
|
7 |
9.96 |
5.1 |
|
8 |
7.94 |
2.35 |
|
9 |
6.44 |
0.72 |
|
10 |
7.59 |
1.64 |
|
11 |
7.88 |
2.24 |
�����������
Figure 5
Graph between CaCO3
and LOI Gravi
The linear regression results were then input into the
hot meal 2023 method on the computer and the result was that the LOI value
displayed in the CaCO3 composition had a value that was almost the same as the
manual/gravity LOI value. So this equation is used by X-ray operators in
analyzing Hotmeal samples every day.
In this special assignment, the
author uses clinker and kiln feed as the main materials in CaCO3 XRD analysis,
manual LOI analysis, and CaCO3 X-Ray analysis packaged in 12 samples with a
content percentage of 80%: 20%. y= 1.1824x-6.9094 which is then input into the
Hotmeal 2023 method to be used as a reference for working on Hotmeal quickly.
Ariadi, D., & Fithriasari, K. (2016). Classification of
Indonesian News Using the Naive Bayesian Classification Method and Support
Vector Machine with Confix Stripping Stemmer. ITS Science and Arts Journal,
4 (2).
Brata, J., & Soediantono, D. (2022). Total quality
manufacturing (TQM) and recommendations for its application in the defence
industry: A literature review. International Journal of Social and
Management Studies, 3 (3), 50�62.
Darmalaksana, W. (2020). Writing Class Experience Research
Formula. Writing Class at UIN Sunan Gunung Djati Bandung, 1.
Effendi, SRNN (2018). Analysis Of The Inventory Of Raw
Materials For Cement Production At PT. Tonasa Pangkep Cement. Journal of
Balanced Economics, Faculty of Economics and Business, 14 (1).
Fajri, A. (2021). The Effect of Using Electric Power on
Cement Residue in the PT Finish Mill Process. Semen Tonasa Unit Tonasa V= The
Effect of Electric Power Usage on Cement Residue in the Finish Mill Process of
PT. Semen Tonasa Unit Tonasa V. Hasanuddin University.
Fauzi, AA, Budi, H., Mulyanto, Dulame, IM, Pramuditha, P.,
& Sudipa, IGI (2023). Utilization of Information Technology in Various
Sectors During Society 5.0. PT. Sonpedia Publishing Indonesia.
Fitri Mulyani, NH (2021). Analysis of the development of
science and technology (IPTEK) in education. Journal of Education and
Counseling. Journal of Education and Counseling (JPDK), 3 (1),
101�109.
Lasdi, L. (2017). Balanced Scorecard as a Framework for
Comprehensive Company Performance Measurement in the Global Business
Environment. Widya Journal of Management & Accounting, 2 (2).
Purba, N., Yahya, M., & Nurbaiti, N. (2021). Industrial
revolution 4.0: The role of technology in the existence of business control and
its implementation. Journal of Business Behavior and Strategy, 9 (2),
91�98.
Radiansyah, D. (2018). The Influence of Technological
Developments on Islamic Youth (Case Study in Citeureup Village, Sukapada
Village). JAQFI: Journal of Islamic Aqidah and Philosophy, 3 (2),
76�103.
Rasul, HA (2017). Marketing Strategy to Increase Product
Sales Volume. Journal of Business and Entrepreneurship, 6 (1).
Sahdiah, H., & Kurniawan, R. (2023). Optimization of
Acceleration Voltage in Scanning Electron Microscope�Energy Dispersive X-Ray
Spectroscopy (SEM-EDX) for Observing the Morphology of Biological Samples. Journal
of Science and Science Education, 6 (2), 117�123.
Sentosa, L., Subagio, BS, Rahman, H., & Yamin, RA (2018).
Activation of Natural Zeolite from Bayah with Acids and Bases as Warm Mixed
Asphalt Additives. Journal of Civil Engineering, 25 (3).
SETIOSOSARI, EKA (2018). Study of the Use of Liquid Smoke
as a Calcium Carbonate (CaCO3) Scale Inhibitor Using the Seeded Experiment
Method.
Soamole, A. M. (2022). Analysis of the Implementation of the
ISO 9001 Quality Management System at PT. Tonasa Cement. Flyover Journal,
2 (1), 1�9.
Zahidin, A., & Rubianto, L. (2020). Calculation Of Mass
Balance, Heat Balance And Efficiency In Rotary Kiln Work Unit RKC 3 PT Semen
Indonesia (Persero) Tbk. Distillates: Journal of Separation Technology, 6
(2), 309�315.